Home brain stimulation devices are legal to buy, but evidence that they work is limited. As of November 2026, the Flow headset and the Alpha-Stim device are sold in the UK without a prescription and the Sooma headset is sold in several EU countries, while trials of home tDCS for depression give mixed results with a small average benefit and NICE rates the evidence for Alpha-Stim in anxiety as low quality.
This page is about buying and using a device at home: what each one is, what it costs, who can sell it, whether the NHS offers it and when a doctor should be involved. For the full evidence review of tDCS as a method, see tDCS for depression: the honest evidence picture.
This page is general information, not medical advice and not a diagnosis. Do not stop or change a medicine or other treatment without talking to your prescriber.
If you are struggling right now. If you are thinking about ending your life or cannot keep yourself safe, do not wait for an appointment. In the UK, call 999 or Samaritans on 116 123, free and around the clock. In the EU, call 112; in many European countries 116 123 also connects to an emotional support line.
Two different technologies
People often group these devices together, but they work in different ways and have different evidence behind them.
| tDCS headsets (Flow, Sooma) | CES device (Alpha-Stim AID) | |
|---|---|---|
| Current | Direct current, about 2 mA | Alternating microcurrent, up to 0.5 mA (500 microamps) |
| Where it goes | Two saline-soaked pads on the forehead, over the prefrontal cortex | Two clips on the earlobes |
| Session | About 20 to 30 minutes, daily or near-daily for several weeks | 20, 40 or 60 minutes, daily, every other day or as needed |
| Marketed for | Depression | Anxiety, insomnia, depression and pain (UK) |
| UK class | Class IIa medical device | Class IIa medical device |
tDCS (transcranial direct current stimulation) passes a weak, steady current between two electrodes on the scalp. For depression, the positive electrode sits over the left dorsolateral prefrontal cortex, an area behind the forehead. The current is too weak to fire neurons directly; the working theory is that it slightly shifts how excitable that area is (our tDCS evidence page).
CES (cranial electrotherapy stimulation) sends a very small alternating current through clips on the earlobes. The Alpha-Stim AID manual gives a range of 0 to 500 microamps at a low pulse rate, with 20, 40 or 60 minute timers (Alpha-Stim AID manual, 2025). How CES might affect mood or anxiety is not well established.
The devices one by one
Flow FL-100 (tDCS headset)
Flow Neuroscience (Malmö, Sweden) sells the FL-100 headset with an app that includes behavioural therapy courses and progress tracking. The EU and UK instructions describe it as a Class IIa medical device to treat mild to severe major depressive disorder in adults aged 18 and over, used at home (FL-100 EU/UK instructions, version 20, July 2026). Flow's declaration of conformity for the FL-100 refers to the earlier EU Medical Device Directive (93/42/EEC). In the UK it carries a UKCA certificate from BSI (BSI certificate UKCA 776047). Flow states that in the UK it "is available for direct purchase, no prescription needed" (Flow NHS page).
In the US the same headset is prescription-only. The FDA approved it in December 2025 (PMA P230024) "for the treatment of moderate to severe major depressive disorder in the current episode, either as monotherapy or as an adjunctive treatment, in patients 18 years and older who are not considered treatment refractory to medication" (FDA PMA database). In other words, the US approval does not cover people whose depression has not responded to medication.
Sooma tDCS
Sooma Medical (Helsinki, Finland) describes its system as EU MDR-certified, a CE Class IIa device for depression and chronic pain, used in a clinic or at home under remote supervision (Sooma). There are two routes. Clinics can own the devices and monitor patients through a clinician portal. Sooma also sells a "Home Therapy" version directly to people in some EU countries: a fixed programme of one 30-minute session a day, with no clinician monitoring included (Sooma: two ways to get started). Its online shop states that no prescription is needed within the EU. On 21 November 2026 the shop delivered to Austria, Finland, France, Germany, Ireland, Luxembourg and Sweden, not the UK (Sooma shop). We did not find a UK clinic offering Sooma.
Alpha-Stim AID (CES)
Alpha-Stim is made by Electromedical Products International in Texas. In Europe it is a Class IIa medical device (Alpha-Stim AID EU manual, 2018). The UK site sells it directly for adults aged 22 and over for anxiety, insomnia, depression and pain (Alpha-Stim UK). The manual says that outside the US it is available without a prescription; in the US federal law restricts it to sale by or on the order of a licensed healthcare practitioner (Alpha-Stim AID manual, 2025).
The US rules treat CES for depression differently from CES for anxiety. In December 2019 the FDA placed CES devices for anxiety or insomnia in class II and kept CES for depression in class III, which requires premarket approval (Federal Register, 20 December 2019; 21 CFR 882.5800). The current US manual is titled for anxiety and insomnia only. The UK site still lists depression and pain.
PsyAccess keeps a list of the devices we have checked against the EU register: the maker, the intended purpose from the maker's own document, the device class, the countries where it is on the market and how to get it. Makers do not pay to be listed.
What the research shows, and where it stops
Home tDCS for depression
This is a short summary of our physician-reviewed tDCS evidence page. Results are mixed, and the average benefit appears small.
- A six-week German trial of tDCS added to an SSRI (DepressionDC, 160 adults) found no added benefit: the difference on the MADRS depression scale was 0.3 points (95% CI -2.4 to 2.9) (Burkhardt et al., Lancet 2023).
- An unsupervised home-use trial in Brazil (210 adults) found no advantage over sham at six weeks (Borrione et al., JAMA Psychiatry 2024).
- A ten-week home trial with remote video supervision in the UK and US (174 adults, using a Flow headset) found a modest benefit: 2.27 points more improvement than sham on the 17-item Hamilton scale (95% CI 0.51 to 4.01) (Woodham et al., Nature Medicine 2024).
- A 2026 analysis of individual data from 18 trials (1,246 people) found a small average symptom benefit, and larger trials showed smaller differences between active and sham treatment (Razza et al., British Journal of Psychiatry 2026).
The trials differed in design and length, so it is not possible to say whether supervision explains the different results. NICE allows tDCS for depression only with special arrangements for clinical governance, consent and audit or research (NICE HTG382). NICE published a briefing on the Flow headset in November 2023 and withdrew it in May 2024 because the evidence had changed (NICE MIB324).
CES and Alpha-Stim
The evidence for CES is older, smaller and of lower quality than the evidence for tDCS.
- Depression. A 2014 Cochrane review looked for randomised trials of CES in acute depression and found none that met its standards; it concluded there were "insufficient methodologically rigorous studies" (Kavirajan et al., Cochrane 2014). The largest sham-controlled trial since then, Alpha-Stim-D, randomised 236 adults with major depression from 25 GP practices in England to an hour a day of active or sham Alpha-Stim for eight weeks. Both groups improved by a similar amount (about 6 points on the Hamilton depression scale at 16 weeks, p=0.46), and the authors concluded the device was safe and acceptable "but no more clinically effective than sham" (Morriss et al., Lancet Psychiatry 2023). The trial was funded by the NIHR and the manufacturer.
- Anxiety. A 2018 systematic review in the Annals of Internal Medicine (26 randomised trials) found "low-strength evidence" of a possible modest benefit in people with anxiety and depression, from four trials done more than 40 years ago and one from 2014. It found the evidence insufficient for depression alone, insomnia and pain, and rated all trials at high risk of bias because blinding was inadequate (Shekelle et al., Annals of Internal Medicine 2018). NICE says Alpha-Stim AID "shows promise for managing anxiety disorders. However, there is not enough good-quality evidence to support the case for routine adoption". It describes the evidence as low quality, with most trials lasting only five or six weeks and only one using the current model (NICE HTG570).
- Newer pooled analyses. Meta-analyses from 2022 to 2025 report a moderate effect on anxiety and a small effect on depression, but they pool small, often older trials (Ching et al., Frontiers in Psychiatry 2022; Liu et al., Journal of Affective Disorders 2025). They do not settle the questions NICE raised, and NICE has asked for a trial comparing Alpha-Stim with CBT, medication or both.
- NHS service evaluation. A study of 161 people waiting for CBT in two English services used Alpha-Stim for 6 to 12 weeks and reported fewer anxiety symptoms. It had no control group, and only 30% of participants completed every assessment (Morriss et al., Journal of Affective Disorders 2019).
The UK seller describes Alpha-Stim as clinically proven to treat anxiety, insomnia, depression and pain. The independent reviews and the NHS-based depression trial above do not reach that conclusion.
UK prices
Prices as listed on the sellers' own websites, checked 21 November 2026. Prices change, so check before relying on them.
| Device | Buy | Rent or hire | Returns |
|---|---|---|---|
| Flow FL-100 | GBP 399 (includes pads for the first 5 weeks) | GBP 79 a month, cancel anytime | 30-day money-back guarantee |
| Sooma Home Therapy | Not sold in the UK; EUR 599 in the EU, or 6 payments of EUR 129 | Not offered | 30-day money-back guarantee |
| Alpha-Stim AID | GBP 599 including VAT | GBP 30 for one week, GBP 108 for four weeks | 60-day money-back on a first purchase, with conditions |
Sources: Flow shop, Sooma shop, Alpha-Stim UK store and terms.
On 21 November 2026 Flow's shop advertised a 20% Black Friday discount (around GBP 319 to buy), with no end date shown. Flow renters can switch to owning the headset (GBP 320 within 30 days, or GBP 199 after three months). Alpha-Stim also sells a refurbished AID for GBP 538.80 and replacement ear clips for GBP 36. Both Flow and Alpha-Stim say VAT relief can apply for people with a diagnosed condition.
Flow pads are consumables: buying includes pads for the first five weeks, and the rental includes refill deliveries. Count the full cost over the length of a course, not only the headline price.
Prescription and regulation at a glance
| UK | EU | US | |
|---|---|---|---|
| Flow FL-100 | Sold directly, no prescription | Sold directly in several countries, CE-marked | Prescription only (FDA approval, December 2025) |
| Sooma | No UK shop found | Through clinics, or direct Home Therapy in some countries without prescription | Investigational only |
| Alpha-Stim AID | Sold directly, no prescription | Sold without prescription, CE Class IIa | Prescription only; CES for depression needs premarket approval |
Being sold without a prescription means you can buy it; it does not mean a clinician has judged it right for you.
Can you get a device through the NHS?
There is no national NHS offer for either device. A few trusts in England have offered Flow in pilots or selected services, so availability depends on your local trust and on which of its services you are under.
- Northamptonshire Healthcare NHS Foundation Trust offered Flow in community mental health teams, the perinatal mental health service and the maternal loss psychology service from January 2024, as an evaluated project (NHFT, January 2024). In May 2024 it said the perinatal and maternal loss services offer it "by clinician referral only" (NHFT, May 2024).
- Leicestershire Partnership NHS Trust piloted Flow in its mental health crisis service for around 80 patients from October 2024 and said it could not offer the headsets to patients of its other services or in primary care (LPT, October 2024).
- Cumbria, Northumberland, Tyne and Wear NHS Foundation Trust published a patient leaflet in July 2026 saying it is testing Flow within the trust, with a consent form for the study (CNTW).
- The NHS Innovation Accelerator, which lists Flow among its 2026 fellows, also names West London NHS Trust, Norfolk and Suffolk NHS Foundation Trust, Rotherham Doncaster and South Humber NHS Foundation Trust and NHS Practitioner Health, and a mother and baby unit pilot at South London and Maudsley (NHS Innovation Accelerator). We did not find these trusts' own pages describing the service.
For Alpha-Stim, Health Innovation East Midlands ran a demonstrator with a GP federation in Northamptonshire, which it says continues to offer the device (Health Innovation East Midlands). NICE does not recommend routine NHS use (NICE HTG570).
If you are under an NHS mental health team, you can ask whether your trust offers a device as part of a service or study.
Outside the UK, tDCS is mainly a clinic treatment rather than a home device paid for by public systems. In Finland, the psychiatric neuromodulation unit at Turku University Hospital (Tyks) lists tDCS among its treatments (Tyks). In Estonia, the Health Insurance Fund's list of services includes a tDCS session under code 7604 (Tervisekassa service list). The provider directory shows which listed services offer tDCS and in what setting.
Safety: who should not use them
Flow FL-100. Under the instructions, the headset must not be used if you have open wounds, broken or damaged skin where the pads sit, or metallic skull reconstruction where the pads sit (sections 5 and 6 of the FL-100 instructions). The instructions ask you to discuss it with your clinician first if you are or may be pregnant, have thoughts of suicide or self-harm, have metal in or around the brain (such as clips, plates, cochlear implants or a brain pacemaker), have epilepsy or seizures, or have a chronic skin condition of the forehead or scalp. The EU and UK instructions state that the headset is not indicated for bipolar disorder, and list suicidal thoughts as a rare and hypomania as a very rare reported reaction (FL-100 EU/UK instructions). Skin redness, itching, stinging and headache are common. Poor pad contact may cause burns.
Alpha-Stim AID. It must not be used with an implanted pacemaker or an implanted or wearable defibrillator. Safety in pregnancy has not been established, and you should not drive or operate machinery during treatment, and in some cases for several hours after. Reported side effects include dizziness, headache and skin irritation at the ear clips (Alpha-Stim UK; manual). NICE describes it as a low-risk device with mild side effects (NICE HTG570).
When to involve a doctor
Talk to your GP or treating team before you start, and keep your current treatment unless your prescriber advises otherwise. Both Flow's instructions and Alpha-Stim's labelling say users should be monitored for signs of worsening.
Contact your treating team promptly, and stop using the device until you have spoken to them, if:
- your mood gets worse, or you start having thoughts of suicide or self-harm;
- you notice signs of a high mood: needing much less sleep, racing thoughts, unusual energy or taking risks you normally would not;
- you have a seizure, a fainting episode, or a headache or skin burn that does not settle.
If you are at immediate risk of harming yourself, call 999 (112 in the EU) or Samaritans on 116 123.
Agree a plan before you start. Agree with your clinician how you will track symptoms, when you will review progress and what you will do if your mood worsens. One simple way to track depression is the PHQ-9 questionnaire, repeated every two to four weeks and brought to appointments. Our PHQ-9 retest tool keeps your scores on your own device so you can show the trend to your doctor. A score is not a diagnosis, and a change in score is something to discuss, not a verdict on the device.
Whether a home device makes sense for you depends on your diagnosis, your treatment so far and your own risks. This is a conversation to have with your doctor.
Questions to ask before using a device
Is this meant to be added to my current treatment, and who will review it? What change would tell us it is helping, and after how long do we stop if it is not? Is there anything in my history (epilepsy, implants, pregnancy, bipolar disorder, skin conditions) that rules it out? Does my local NHS trust offer it within a service or study? What will the full course cost, including pads or ear clips?
For other options after standard treatment, see the beyond antidepressants overview and rTMS for depression in Europe.
Frequently asked questions
Do home brain stimulation devices work for depression?
For home tDCS such as Flow, trial results are mixed and the average benefit appears small; a remotely supervised UK and US trial found a modest benefit, while an unsupervised home trial and an SSRI add-on trial found none. For CES such as Alpha-Stim, a 2023 NHS-based trial in major depression found it no better than a sham device.
Does Alpha-Stim work for anxiety?
Some small trials suggest a modest benefit, but the evidence is low quality. NICE says Alpha-Stim "shows promise" for anxiety disorders but that there is not enough good-quality evidence to support routine use, and it has asked for a trial against CBT or medication.
What is the difference between Flow and Alpha-Stim?
Flow is a tDCS headset: a direct current of about 2 milliamps through pads on the forehead, sold for depression. Alpha-Stim is a CES device: an alternating microcurrent of up to 0.5 milliamps through clips on the earlobes, sold in the UK for anxiety, insomnia, depression and pain.
Do I need a prescription for Flow or Alpha-Stim in the UK?
No. Both are sold directly in the UK without a prescription. In the US both are prescription-only.
Is the Flow headset available on the NHS?
Not as a national offer. Some NHS trusts in England have offered Flow in pilots or in selected services, often by clinician referral, so availability depends on your local trust and on the service you are under.
How much do Flow and Alpha-Stim cost in the UK?
On 21 November 2026, Flow listed GBP 399 to buy or GBP 79 a month to rent, with a temporary Black Friday discount. Alpha-Stim AID listed GBP 599 to buy or GBP 30 a week to hire.
Sources
All pages checked 21 November 2026.
- Burkhardt G, et al. DepressionDC. The Lancet 2023;402:545-554. pubmed.ncbi.nlm.nih.gov/37414064
- Borrione L, et al. Home-use tDCS for a major depressive episode. JAMA Psychiatry 2024;81(4):329-337. pubmed.ncbi.nlm.nih.gov/38170541
- Woodham RD, et al. Home-based tDCS treatment for major depressive disorder. Nature Medicine 2024. doi.org/10.1038/s41591-024-03305-y
- Razza LB, et al. Individual patient data meta-analysis. British Journal of Psychiatry 2026. pubmed.ncbi.nlm.nih.gov/41498282
- Kavirajan HC, Lueck K, Chuang K. Alternating current cranial electrotherapy stimulation (CES) for depression. Cochrane Database of Systematic Reviews 2014, CD010521. cochrane.org/evidence/CD010521
- Shekelle PG, et al. Benefits and harms of cranial electrical stimulation for chronic painful conditions, depression, anxiety, and insomnia: a systematic review. Annals of Internal Medicine 2018;168(6):414-421. pubmed.ncbi.nlm.nih.gov/29435567
- Morriss R, et al. Alpha-Stim-D: active versus sham Alpha-Stim AID in major depression in primary care in England. Lancet Psychiatry 2023;10(3):172-183. pubmed.ncbi.nlm.nih.gov/36724796
- Morriss R, et al. Alpha-Stim CES in treatment-seeking patients with generalised anxiety disorder. Journal of Affective Disorders 2019;253:426-437. pubmed.ncbi.nlm.nih.gov/31103808
- Ching PY, et al. CES for anxiety: meta-analysis. Frontiers in Psychiatry 2022;13:899040. pubmed.ncbi.nlm.nih.gov/35757229
- Liu SY, et al. CES for depression: meta-analysis. Journal of Affective Disorders 2025;382:488-497. pubmed.ncbi.nlm.nih.gov/40286924
- NICE HealthTech guidance HTG570, Alpha-Stim AID for anxiety disorders (first published March 2021 as MTG56). nice.org.uk/guidance/htg570
- NICE HealthTech guidance HTG382, transcranial direct current stimulation (tDCS) for depression (2015, formerly IPG530). nice.org.uk/guidance/htg382
- NICE medtech innovation briefing MIB324, Flow (published November 2023, withdrawn May 2024). nice.org.uk/advice/mib324
- US FDA PMA P230024, Flow FL-100, decision 8 December 2025: database entry and instructions for use (accessdata.fda.gov/cdrh_docs/pdf23/P230024C.pdf).
- US Federal Register 84 FR 70003, 20 December 2019, CES reclassification (govinfo.gov); 21 CFR 882.5800.
- Flow Neuroscience: shop (flowneuroscience.com/shop), NHS page (flowneuroscience.com/nhs-flow-pilot), EU/UK instructions for use version 20, July 2026.
- BSI certificate UKCA 776047 (Flow Neuroscience AB).
- Sooma Medical: soomamedical.com, soomamedical.com/two-ways-to-get-started-with-sooma-tdcs, shop.soomamedical.com.
- Alpha-Stim UK: alpha-stim.co.uk/store, alpha-stim.co.uk/risk, alpha-stim.co.uk/terms-and-conditions; Alpha-Stim AID owner's manual IFU-CUS-002 version 3 (2025); EU manual REV K (2018).
- Northamptonshire Healthcare NHS Foundation Trust news, 16 January 2024 and 15 May 2024 (nhft.nhs.uk).
- Leicestershire Partnership NHS Trust news, 28 October 2024 (leicspart.nhs.uk).
- Cumbria, Northumberland, Tyne and Wear NHS Foundation Trust patient leaflet PIC/963/0726, July 2026 (cntw.nhs.uk).
- NHS Innovation Accelerator, Flow Neuroscience fellow page (nhsaccelerator.com).
- Health Innovation East Midlands, cranial electrotherapy stimulation for anxiety project page (healthinnovation-em.org.uk).
- Tyks psychiatric neuromodulation unit (tyks.fi); Tervisekassa list of health services, version 52.0.0, code 7604.